Search
Menu
Spectrogon US - Optical Filters 2024 LB

Improving Capillary Electrophoresis the Photonic Way

Facebook X LinkedIn Email
Novel techniques using lasers, LEDs and fiber optics aid protein separation and analysis.

Lynn M. Savage, News Editor

As a tool for divining all that is knowable about proteins, capillary electrophoresis has taken a strong grip on investigators. The technique offers scientists from oncologists to pharmacologists speedy and sensitive separation of proteins — either inside narrow tubes or within the etched walls of a microfluidic chip — and a number of methods have been developed to detect and analyze the proteins within a sample, including laser-induced fluorescence.

But there is always room for improvement. Researchers continuously seek ways to reduce both the size of bulky equipment and the costs of analytical systems and, most importantly, to improve the speed and sensitivity of the process so that protein analysis can reveal more information.

Optical injection

Traditional sample injection in capillary electrophoresis involves physical manipulation of the separation capillary into and out of vials containing buffer or sample, making the process slow and difficult to reproduce. These limitations have led to the development of so-called gating techniques that enable injection into a continuously flowing solution. Gated injection — whether by mechanical or optical means — segments an analyte within a buffer into easy-to-analyze chunks that then pass within view of a detector, such as a laser beam.

In optically gated capillary electrophoresis, the sample to be studied is tagged with a fluorescent dye, added to a buffer solution and continuously streamed under the high-voltage electrical field. A high-power laser photobleaches the analyte as it passes by the beam’s target area. By intermittently blocking the laser beam, a select amount of actively fluorescent analyte can be detected by a second laser farther down-stream.

Although this method provides faster processing times than traditional, nongating methods, the photobleaching step rarely is more than 90 percent efficient. Inefficient bleaching leads to significant background fluorescence at the detection stage and, thereby, to lower sensitivity.

In the June 1 issue of Analytical Chemistry, Craig A. Aspinwall and his colleagues at the University of Arizona in Tucson describe an alternative approach that puts a twist into standard optical gating.

The researchers tagged analytes with a fluorophore that was “caged” via covalent attachment of a pair of nitrobenzyl groups to a molecule of fluorescein to inhibit fluorescence. When added to buffer running through a capillary electrophoresis system, the analyte was effectively invisible to the filtered 488-nm detection beam from an argon-ion laser made by Coherent Inc. of Santa Clara, Calif.


Figure 1.
A schematic of a photolytic optical injection method for capillary electrophoresis depicts laser beam orientation.


However, when the cages were removed via photolysis with the 351- to 364-nm lines split off from the same beam, the fluorescence was reactivated and the analytes became detectable. The investigators directed the focus of the photolytic beam onto the capillary for periods of 5 to 20 ms using a mechanical shutter and collected fluorescence using a 20x, 0.4-NA objective from Melles Griot and a photomultiplier tube from Hamamatsu Photonics in Bridgewater, N.J. (Figure 1).

According to Aspinwall, using photolysis to activate fluorescence reduced background noise by as much as 10,000 times, with a corresponding increase in sensitivity that should enable detection of neuropeptides and other proteins that are difficult to elucidate (Figure 2).


Figure 2.
During the separation of amino acids labeled with caged fluorescein using a photolytic optical gating technique, consecutive online injections were performed on a sample mixture containing 50 nM labeled arginine (Arg), glutamic acid (Glu) and aspartic acid (Asp). Lines in upper trace indicate times of injection. E = 1.77 kV/cm; injection time = 20 ms.


After uncaging, the analytes separate as they travel from the site of the gating beam to that of the detection beam. In the setup, the discrete photolytic and detection laser lines were spaced 1.4 cm apart along the capillary. The investigators would like to shrink that distance, which thus far is limited by the physical constraints of the filters and other optical components that they employed. “Shorter separation paths yield higher speed,” Aspinwall said. “Ideally, one would like to use subcentimeter separations.”

The researchers are focusing on three approaches to improve their technique. They are exploring the use of LEDs to provide the UV and blue wavelengths now generated by the laser, developing caged fluorescent labels that have faster reaction dynamics and integrating the system into a microfluidic platform.

Getting the LED in


Aspinwall’s group is, by no means, the only one looking to LEDs to replace lasers in capillary electrophoresis systems. Given the ongoing drive to reduce the size and cost of protein analysis, lasers — even solid-state diode devices — are bulky and expensive. Antonio Segura-Carretero of the University of Granada in Spain, said LEDs are “an exceptionally attractive stable excitation light source for luminescence detection in capillary electrophoresis.”

He and his colleagues in Alberto Fernández-Gutiérrez’s research group in the department of analytical chemistry have explored methods to improve the way LEDs can be used to excite fluorescent labels at the detection stage.

The scientists noted that most efforts to swap LEDs for lasers involve using microscopes or plastic lenses to focus the incoherent light down to the capillary as well as to collect the resulting fluorescence. The LED-based systems used thus far also tend to be made of noncommercial components and require an expert hand to assemble the complicated apparatus.

Using an optical fiber instead of a complex lens system to bring the LED output to the capillary and the fluorescence emission to the detector improves simplicity while reducing cost but, according to Segura, previous attempts to do so resulted in difficulty aligning the capillary and fiber. The researchers report in the May issue of Electrophoresis that they have devised a simple system that addresses these disadvantages.

They used an LED-based pulsed light source from Ocean Optics BV in Duiven, the Netherlands, that emits light at a maximum of 470 nm. Using a customized detection cell, they perpendicularly aligned the capillary with one end of a bifurcated fiber optic probe from Avantes BV in Eerbeenk, the Netherlands, and attached the split ends of the fiber to the light source and to a spectrometer also from Avantes. The spectrometer included a linear-array CCD detection chip with 2048 pixels (Figure 3), a 600 lines/millimeter grating and a 500-μm slit, which together provided a resolution of 10 μm FWHM.


Figure 3.
To test the ability of fiber optic-linked LEDs to replace lasers for exciting fluorescence during capillary electrophoresis, researchers at the University of Granada in Spain used a pulsed light source outfitted with an LED that emits at 470 nm, a bifurcated fiber optic probe and a spectrometer with a linear-array CCD detection chip with 2048 pixels. Courtesy of Antonio Segura-Carretero.


The design of the detection cell enabled the investigators to adjust the distance between the end of the fiber and the outer edge of the capillary, and they found that the narrower the gap, the higher the sensitivity: At a distance of 2 mm, the limit of detection was 28 μM; at 0.05 mm, it was 2.6 μM.

In tests involving various naturally fluorescent, nonfluorescent and fluorescently tagged compounds, the researchers found that the system provided adequate detection capability, albeit with detection limits not as low as can be achieved with laser-induced fluorescence. They believe that the detection limitations of the LED-based system can be improved by using LEDs brighter than the 50-μW device that they used, a more sensitive CCD detector chip, a more efficient fiber optic or a combination of the three.

Sample preconcentration

Among the techniques that have been developed to improve capillary electrophoresis are field-amplified stacking injection (FASI) and microfluidics.

In FASI, water or another low-conductivity solution is introduced into a buffer-filled capillary, followed by some amount of sample solution. When a reverse-polarity electric field is applied to the capillary, the analyte’s anions accumulate, or stack, at the boundary between the buffer and the sample solution while the plug of water is pumped out of the capillary entirely by the electro-osmotic flow. Stacking the ions in this manner concentrates the analyte ions as they pass the detector, increasing the sensitivity of measurements without significantly losing resolution.

Meadowlark Optics - Wave Plates 6/24 MR 2024

Shrinking capillary electrophoresis experiments to the microfluidic scale reduces sample amounts, processing time and, ultimately, costs. However, performing experiments on a microchip — especially when trying to integrate FASI techniques — is complicated by the open system of valveless channels that are used in place of a single capillary.

Now, Maojun Gong, William R. Heineman and Patrick A. Limbach of the University of Cincinnati’s department of chemistry, and Kenneth R. Wehmeyer and Francisco Arias of Procter & Gamble Pharmaceuticals in Mason, Ohio, have described a method to bring FASI to microchip capillary electrophoresis.

They report in the June 1 issue of Analytical Chemistry that they used the simplest possible microchannel design — a regular cross in which four channels meet at an intersection at 90° — to demonstrate that the common variations in velocity within the bulk solution could be used to facilitate the FASI technique.


Figure 4.
To demonstrate the feasibility of field-amplified stacking injection in a microfluidic chip, investigators took advantage of the mismatched flows of buffer and sample within a simple, cross-shaped system. By manipulating the voltage configuration among the four points of the cross, they loaded sample solution system (A), created sample buffer plugs (B), stacked analyte ions along the boundaries between low- and high-concentration buffers that accumulate in the long arm (C), and positioned the stacked ions at the point of detection (D). Courtesy of the American Chemical Society.


At the ends of channels etched into a cruciform upon a glass chip, the researchers attached reservoirs to hold the sample, the buffer and the waste from each (Figure 4). An applied voltage across the short arm of the cross pumped low-conductivity sample solution from its reservoir through the intersection and into its designated waste receptacle. To inject a low-conductivity sample buffer plug into the separation channel to serve as the FASI medium, they switched the voltage to run between the sample reservoir and the buffer waste receptacle. This drove the sample into all of the channels except the separation channel, in which only a low-conductivity sample buffer plug was introduced, while the anionic analytes were filtered out by the high electric field there.

According to Gong, who led the experiment, in a traditional sealed capillary, the velocity of the bulk fluid is the average of the faster moving low-conductivity material and the slower high-conductivity solution. In the open channels of a microchip, the high-conductivity material has more freedom to move, and it acts to induce solution flow within the branches, enabling the next step.

To stack the ions, the investigators reversed the polarity of the prior step, applying a voltage between the sample and buffer waste receptacles. This caused the bulk fluid to flow toward the sample waste receptacle from each of the channels, through the intersection. During this migration, the analyte’s anions accumulated at the boundary between the low- and high-conductivity solutions.

Just before the backward-running plugs reached the intersection, Gong switched the voltage again, turning the plug away from the intersection. This separated the stacked analyte ions into zones that then passed by a detection system composed of a photomultiplier tube from Hamamatsu Corp. of Bridgewater, N.J., a CCD camera from Dage-MTI Inc. of Michigan City, Ind., an inverted microscope from Nikon Corp. of Melville, N.Y., and a 150-W xenon flashlamp emitting at ~480 nm. They used neutral and anionic fluorescent dyes to image the FASI process inside the chip (Figure 5).


Figure 5.
Separately using two fluorophores (A, B), the researchers imaged the field-amplified stacking injection process depicted in Figure 4C. The arrows indicate the bulk flow direction. SW = sample waste, SR = sample reservoir, BW = buffer waste, BR = buffer reservoir. Courtesy of the American Chemical Society.


The researchers found that the length of the sample buffer plug influenced the level of detection enhancement that they could achieve. By varying the duration of the plug-injection step, they manipulated the plug length. If the plug was too short or too long, signal enhancement was reduced; ultimately, they found that an injection time of 15 s was optimal. In experiments with fluorescein-5-isothiocyanate, fluorescein disodium salt and 5-carboxyfluorescein, they recorded detection enhancements of 94-, 108- and 160-fold, respectively, relative to the commonly used pinch injection method.

They note that the technique, thus far, works only with anions, but that minor modifications would enable the detection of cationic analytes as well. Gong also said that there would be value in testing the technique in more complex microchip designs, because advanced chips have additional channels that enable such processes as mixing and chemical reactions.

Coating capillaries

No matter how successfully you shrink a capillary system or optimize the sample concentration, analyzing proteins via electrophoresis is problematic: Because of ionic interactions, hydrogen bonding and van der Waals interactions between the sample and the glass or fused-silica material, proteins tend to adhere to the inside wall of the capillary or microchannel, extending the sample volume and thereby reducing sensitivity.

Coating capillaries to make them less prone to protein adsorption is common, but existing coatings do not fulfill the three main requirements for optimizing capillary electrophoresis: high resistance to protein adsorption, the ability to limit electro-osmotic flow, and high stability, which is necessary for compatibility with online mass spectrometry techniques.

At West Virginia University in Morgantown, Aaron T. Timperman and his colleagues have developed a coating composed of polyethylene glycol (PEG) and silane that addresses each of these requirements.

The researchers report in the July 1 issue of Analytical Chemistry that they used PEG because it provides excellent resistance to protein adhesion. To investigate the optimal combination of PEG and silane, they formed two coatings: one composed of silane terminated with a PEG with a molecular weight of 750, and another with a molecular weight of 4000 to 5000. They also concocted a third coating, composed of a mixture of the first two.

Using a capillary electrophoresis system from Beckman-Coulter Inc. of Fullerton, Calif., featuring a 488-nm argon-ion laser for fluorescence excitation and a photodiode array for detection, the scientists tested the coatings on fused-silica capillaries. They also tested them on a glass microfluidic chip.

After the coating step, they exposed the channels and capillaries to various proteins — bovine serum albumin, for example — that were tagged with fluorescein isothiocyanate. They then measured the fluorescence intensity of the adsorbed protein-fluorophore compounds using an inverted microscope from Nikon, a 4x objective, a 530 ±30-nm emission bandpass filter and a CCD camera from Roper Scientific in Tucson, Ariz.

They found that the coatings composed of the larger PEG molecules and those composed of a mixture of large and small PEG particles most efficiently reduced electro-osmotic flow, suppressing it by >90 percent. Using short PEG molecules alone achieved 65 percent reduction in electro-osmotic flow.

The coatings were stable, with those on the glass microchannels dropping in efficiency to an equivalent flow reduction of ~84 percent after about 30 runs. Coatings on the fused-silica capillaries lasted for more than 200 runs, with the flow reduction dropping to 78 percent.

According to Timperman, the characteristics of the coatings, especially their stability over a large number of runs, makes them suitable for mass spectrometry, and his group foresees testing them in combined capillary electrophoresis/mass spectrometry experiments.

“Using capillary electrophoresis with ultraviolet or other types of optical detection will provide quantitation,” he said, “while mass spectrometry will provide the characterization.

Published: August 2006
Basic ScienceBiophotonicsCoatingsFeaturesMicroscopySensors & Detectorsspectroscopy

We use cookies to improve user experience and analyze our website traffic as stated in our Privacy Policy. By using this website, you agree to the use of cookies unless you have disabled them.