Search
Menu
Hamamatsu Corp. - Mid-Infrared LED 11/24 LB

Photonic Sensors Withstand Large Doses of Radiation

Facebook X LinkedIn Email
Researchers at the National Institute of Standards and Technology (NIST) have identified a class of silicon photonic sensors that could be effective in high-radiation environments because of their ability to withstand high radiation doses. Results from tests performed by the team indicate the durability of such devices under harsh conditions and suggest that these sensors could be customized to measure radiation doses in both industrial applications and clinical radiotherapy.

The researchers exposed two kinds of silicon photonic sensors — silicon ring resonators and Bragg waveguides — to hours of cobalt-60 γ-ray radiation up to 1 megagray (MGy). The Bragg waveguides and ring resonators exposed to varying levels of γ-ray radiation did not show any significant changes in spectral characteristics.

A NIST prototype photonic thermometer. Courtesy of Jennifer Lauren Lee/NIST.
A NIST prototype photonic thermometer. Courtesy of Jennifer Lauren Lee/NIST.

Measurement of device characteristics such as peak center, peak width, force sensitive resistor (FSR), and temperature sensitivity did not reveal any significant dose-dependent changes, indicating that for all devices tested, the characteristic group index, dispersion, and thermo-optic coefficient would remain constant for aggregate doses up to 1 MGy, the maximum absorbed dose delivered in the study.

“We found that oxide-coated silicon photonic devices could withstand radiation exposure up to 1 million gray,” said researcher Ryan Fitzgerald. “It’s the upper limit of what our calibrations customers care about. So the devices can be assumed to work reliably at industrial or medical radiation levels that are hundreds or thousands of times lower.”

One gray represents 1 joule of energy absorbed by 1 kilogram of mass and corresponds to 10,000 chest x-rays. This is roughly what a sensor would receive at a nuclear power plant.

The NIST team’s results suggest that these sensors could be used to track levels of ionizing radiation (i.e., radiation with energy high enough to alter the structure of atoms) used in food irradiation to destroy microbes. The sensors could also be used in medical device sterilization, and could have potential applications in medical imaging and therapy.

Teledyne DALSA - Linea HS2 11/24 MR

The results could also have implications for new medical therapies that employ extremely narrow beams of protons or carbon ions, and for medical sterilization processes that use low-energy beams of electrons. “Our sensors are naturally small and chip-scale,” Fitzgerald said. “Current dosimeters are on the order of millimeters to centimeters, which can give erroneous readings for fields that vary over those dimensions.”

In the next stage of its research, the team will test arrays of sensors simultaneously in identical conditions to see if variations in dose over small distances can be resolved.

“Eventually, we hope to develop chip-scale devices for industrial and medical applications that can determine absorbed dose gradients over distances in the range of micrometers and thus provide unprecedented detail in measurements,” said researcher Nikolai Klimov.

The research was published in Scientific Reports (https://doi.org/10.1038/s41598-018-31286-9).


NIST scientists tested whether radiation impairs the performance of photonic sensors. In this setup, scientists put a photonic sensor under a powerful radiation-producing electron beam. The beam traveled down a shaft and hit a shutter that can be closed to stop the beam or opened to allow the beam through. When the beam hit the chip, it glowed. From the control room next door, physicists could manipulate the beam and monitor the data to ascertain how the chip was performing. Scientists hope to use these chips someday as sensors in areas with high levels of radiation, such as in space or near equipment used for medical imaging and cancer radiation therapy. Courtesy of Jennifer Lauren Lee/NIST and Silent Partner.

Published: November 2018
Research & TechnologyAmericasNISTNational Institute of Standards and TechnologyMaterialsSensors & Detectorssilicon photonicsphotonic sensorssilicon photonic sensorsindustrialOpticsapplied opticsaerospaceenergyBiophotonicsphotonic thermometerBioScanTech Pulse

We use cookies to improve user experience and analyze our website traffic as stated in our Privacy Policy. By using this website, you agree to the use of cookies unless you have disabled them.