Why Photomultipliers Need Amplifiers
Dr. A.G. Wright
Although photomultipliers come with versatile internal amplifiers, certain applications also require external amplification.
Photomultipliers remain the preferred
light detectors for many applications, mainly because they combine a detection area
up to 1500 cm
2 with electron multipliers — internal amplifiers of exceptional
performance. Besides low noise, near-zero offset and low and constant output capacitance,
electron multipliers deliver gain up to 10
8 and bandwidth up to 500 MHz. Given such
performance, why should photomultipliers need additional amplification?
In practice, the photomultiplier is one of many
elements in a detection system. System designers optimize overall sensitivity by
balancing the allocation of gain between the photomultiplier and electronics. Obviously,
this requires operating both elements according to which functions they perform
satisfactorily.
Photomultipliers perform some functions
better than others. External electronic amplifiers help fill in the function and
performance gaps, and they have functions other than amplification, such as impedance-matching,
bandwidth-limiting and filtering, and pulse-shaping.
Figure 1. Overall sensitivity of a measuring system is proportional
to the product of each element’s gain. As illustrated for a scintillation
counter where measurements are displayed on a multichannel analyzer, the combination
g1 g2 g3 g4 will achieve the required sensitivity.
Photomultiplier gain declines unpredictably
during continuous operation. A significant cause is the total charge taken from
the multiplier. External amplification is advisable in applications requiring continuous
anode current of more than 1 μA. Using an amplifier with gain of 100, for
example, permits a corresponding decrease in photomultiplier gain and, thus, an
anode current of only 10 nA.
Peak signal current performance is
also a consideration. Photomultipliers with linearly focused dynodes can deliver
peak currents up to about 100 mA, or equivalently, a peak signal of 5 V into 50
Ω. Other dynode types do not offer the same peak current capability and may necessitate
additional amplification.
All photomultiplier applications fall
into three categories: photon counting, pulsed signal measurement and analog detection.
All three present occasions in which external amplification is useful.
Figure 2. The equivalent circuit for a photomultiplier is a current generator, in parallel
with resistance, Ro, being much greater than 109 Ω and with stray capacitance,
C0, about 5 pF. In this example, the current generator mimics a scintillator with
decay time constant τ. Most applications allow analysis in terms of an equivalent
R and C combination. Without loss of generality, we can take R = R0R/(R0+R) and
C = C1 + C0.
Photon counting
In photon counting, single photoelectrons initiate
photomultiplier output signals. Counters simply tally all of the signals that exceed
a fixed voltage threshold. This requires conversion of the current signal into voltage
and amplification of the voltage signal to about 50 mV, a level compatible with
commercially available discriminators.
Consider a simple resistor, R, connected
to the anode to convert current to voltage. This resistor might see a stray capacitance,
C, of at least 10 pF contributed by the photomultiplier, the socket and voltage
divider board with time constant
τ1 = RC. If a coaxial cable or oscilloscope
probe is added, it could easily add another 20 pF of stray capacitance. We can approximate
the single electron signal at the anode by:
i(t) = (—eg/τ)exp(—t/τ)
Where e is the electronic charge, g
is the gain of the multiplier and
τ is the decay time of the photomultiplier
pulse. For analysis purposes, an exponential decay is a sufficiently accurate representation
of the anode output. The output voltage v
0(t) for such a stimulus is given by:
v0(t) = —egR/(τ1 —τ).{exp(—t/τ)
— exp(—t/τ1)} provided τ ≠ τ1
This equation suggests that increasing
resistance can increase the amplitude, but that also increases
τ1, and if
τ1 is much greater than
τ, the equation shows, v
0(peak) amplitude varies
inversely with stray capacitance, which should be kept small.
If, as is often the case, an application
requires coverage of a wide dynamic counting range, then it requires
τ1 to
be comparable to
τ. Otherwise, the imposed time constant will broaden the
output pulse, leading to overlap and dead-time effects.
Figure 3. In the second equation, stray capacitance, C, is 10 pF; the decay time of the
photomultiplier pulse, τ, is 5 ns; and the gain of the multiplier, g, equals
106 for various values of resistance, R. An oscilloscope can confirm the amplitude
and the shape of the pulses.
A plot of the second equation (Figure
3) illustrates these points and indicates maximum amplitude of only 1.2 mV when
resistance equals 50 Ω. If we are looking for a 50-mV signal, the necessary
photomultiplier gain is about 4 x 10
7. Photomultipliers are capable of operating
at this gain, but at a rate of 10 MHz, the mean anode current will be:
Ia = 1.6 x 10—19 x 4 x 107 x 107 = 64 μA
The photomultiplier will lose gain
over time, so it is preferable to seek additional gain from an electronic amplifier.
A transimpedance amplifier is ideal for this purpose. It converts input current
into output voltage, which is just what’s required.
The second equation also describes
output from the amplifier with feedback resistor. This amplifier configuration isolates
the voltage-generating resistor from stray capacitance associated with the photomultiplier.
A well-laid-out circuit board will limit stray capacitance seen by the resistor
to about 0.1 pF and, provided the operational amplifier is sufficiently fast and
stable, will allow resistor values up to about 10 kΩ. Such values will provide
a peak signal of about 320 mV when photomultiplier gain is 10
6. It is best to mount
the amplifier as close to the anode as possible to minimize input stray capacitance
— always undesirable in fast electronics.
Figure 4. Diagram (a) shows current-to-voltage
conversion at the anode of a photomultiplier. Diagram (b) illustrates a transimpedance
amplifier, which, with its virtual earth input, is ideally suited to current generators
such as photomultipliers. The best amplifiers are still made with discrete components,
although fast and stable operational amplifiers are now more readily available.
In theory, resistance in (b) may be as high as 10 kΩ.
Transimpedance amplifiers offer the
additional advantage of low output impedance, typically less than 50 Ω, making
them suitable for driving subsequent circuitry without loading the resistor. This
isolation is impossible if the resistor is connected directly to the anode, in which
case there is no drive capability. An alternative is to use a voltage amplifier,
available in a nuclear instrument module unit or, ideally, as a board connected
directly to the photomultiplier base. These typically have 50-Ω input impedance
and provide gain from 10 to about 100.
Pulsed applications
Most multiphotoelectron applications are in the
field of scintillation spectroscopy. However, users may adapt the considerations
discussed here to other pulsed applications.
Energetic charged particles can be
detected by measuring the light emitted when they interact with an organic scintillator.
The decay time constant of the light is typically 1 to 2 ns, comparable with the
time resolution of the fastest photomultipliers. The second equation covered this
application, but here the pulses are n times bigger — n being the number of
photoelectrons in each pulse.
High-density inorganic scintillators are used to detect γ rays. NaI(Tl) is the most widely used material, and it produces about 30 photons per keV of γ-ray energy deposited. A γ-ray of 1 MeV will thus generate about 6000 photoelectrons from a bialkali cathode with an exponential decay time of 240 ns.
Scintillation spectroscopy derives
information based on measurement of the rate of events or the size and time of the
occurrence of each pulse. Measuring rate is relatively straightforward, but determining
the number of photoelectrons in each pulse is more complicated. The ability to see
individual photoelectrons in an event decreases with increasing energy. For energies
above a few kiloelectronvolts, for instance, individual photodetections may not
be separated enough to be counted, leading to underestimation of the event’s
energy.
A better way to measure the energy
is to integrate the signals and convert the sum of photodetections into a single
pulse, the area of which is directly related to the number of photodetections in
the event. Connecting the photomultiplier output to a capacitor, C, will integrate
each charge pulse, regardless of whether it overlaps. Also, the capacitor will charge
to a voltage proportional to the total number of photoelectrons generated.
Again, transimpedance amplifiers are
ideal for this purpose. However, in this case, capacitors in the feedback loop perform
the integration. The capacitor must discharge between events and, with NaI(Tl) scintillators,
it is usual to choose parallel resistors, R, so that τ
1 = CR is about 1 μs,
or four time constants. This ensures that the energy determination of the event
encompasses most of the output light and enables count rates up to about 20 kHz.
Measuring the occurrence time of each
pulse is beneficial if based on the arrival of the event’s first photoelectron,
but such low-level detection is not always practical. It is customary to follow
the integrating amplifier by a differentiating one. This terminology may be confusing
because integration followed by differentiation constitutes a null operation, so
why do it? The processes involved, in fact, bear only similarities to differentiation
and integration, but the terminology is well-entrenched.
An amplifier with a capacitor at its
input passes no direct current, so the unipolar pulses (Figure 5a) are converted
to bipolar pulses. The time of zero crossing is independent of pulse amplitude,
which is highly desirable in critical timing applications.
Figure 5. In diagram (a), capacitor
C integrates output from the photomultiplier producing an output v(max) of about
Q/C. Readers can verify that a 1 MeV γ producing 6000 photoelectrons will
generate a 1-V output for C = 100 pF with photomultiplier gain = 105. Diagram (b)
indicates that the “differentiating” amplifier produces bipolar pulses
that cross the time axis at a common point. A zero-crossing discriminator detects
the time of occurrence of this point.
Analog applications
Film scanning, an analog application, attaches
a positive film to a rapidly rotating drum and exposes it to a sharply focused intense
light source. The light source and a photomultiplier, which detects light reflected
from the film, traverse the drum to scan its contents for subsequent digitization
by an analog-to-digital converter. Bandwidth is the key parameter here and must
be optimized according to the speed of the drum and the required pixel size. High-resolution
machines may require bandwidths up to 5 MHz.
Once again, the best way to provide
the required bandwidth is to use a transimpedance amplifier.
An intense light source ensures high
resolution and minimum scan time. A photomultiplier with only six dynodes provides
comparatively low gain, in the 10
3 to 10
4 range. It is important in these applications
to maintain a mean anode current below 1 μA to ensure stable gain. Excursions
up to 100 μA are permissible, but only for short periods of time.
Important points to remember include:
• The photomultiplier is a current
generator that produces a charge pulse. For encoding, it requires conversion to
a voltage.
• The output may contain unwanted
fine structure, and amplifiers achieve pulse smoothing and shaping.
• Amplifiers can eliminate photomultiplier
fatigue by providing a portion of overall system gain.